How two identical Instruments can provide different results.

View HSQ Kit Ordering Information

Even identical HPLC systems from the same manufacturer can produce different chromatographic results. Differences in system sensitivity, extra-column volume, dwell volume, detector performance, and overall instrument condition can affect method transfer, system suitability, limits of detection, and analytical results.

The example below shows actual performance qualification data from two identical HPLC instruments operating in the same GMP laboratory. A stability-indicating method was successfully developed on one system, but when transferred to the second instrument, several peaks were not detected as expected. Using the MICROSOLV HSQ Kit standard qualification protocol, the laboratory identified measurable performance differences between the two instruments despite having the same model, configuration, and similar usage history.

This example demonstrates why instrument characterization is important for method development, method transfer, troubleshooting, and ongoing laboratory quality assurance.

By routinely measuring key system performance characteristics such as sensitivity, dwell volume, extra-column dispersion, injection precision, retention-time precision, and flow-rate accuracy, laboratories can better understand instrument-to-instrument variability and make more informed analytical decisions.

The MICROSOLV HSQ Kit provides a practical and standardized approach for evaluating and documenting HPLC system performance across multiple instruments, departments, and laboratory locations.

Instrument:   Agilent 1100 LC System #9 Agilent 1100 LC System #2
Performance Characteristic: Test Conditions: Performance: Performance:
System Extra-Column Dispersion: Rs test mix extrapolated N 5 uL 3 uL
Gradient Dwell Volume: Dwell Volume: 1.20 mL 1.07 mL
  Delivery Accuracy: All circuits <5% of nominal All circuits <5% of nominal
% Injector Carryover: No Wash Vial: 0% 0%
With Wash Vial: 0% 0%
Injection Precision: 10 replicate inj. 0.07% RSD 0.07% RSD
Ret. Time Precision: 10 replicate inj. 0.21% RSD 0.06% RSD
System Sensitivity: caffeine @ 273nm normalized for column efficiency

peak to peak noise @ 273 nm
Sens. = 363.9 mAU/µg

LOD = 0.032 µg/mL

Noise = 0.04 mAU
Sens. = 400.6 mAU/µg

LOD = 0.058 µg/mL

Noise = 0.077 mAU
Pump Flow Rate Accuracy: 0.5 mL/min 100.3% 100.0%
1.0 mL/min 100.3% 100.3%
5.0 mL/min 99.7% 99.7%
Pump Max Press: BPR, leak check Qual: to 5,000 psi, 345 bar Qual: to 5,000 psi, 345 bar
Column Oven: oven air temp
<5°C of normal
Qual: 20-50°C, <5°C of normal Qual: 20-50°C, <5°C of normal